Plastics Technology Laboratories  
 

Compositional Analysis by Thermogravimetric Analysis (TGA)

ASTM E1131, ISO 11358

Click on the picture for a larger view


Scope:
In a Thermogravimetric Analysis, the percent weight loss of a test sample is recorded while the sample is being heated at a uniform rate in an appropriate environment. The loss in weight over specific temperature ranges provides an indication of the composition of the sample, including volatiles and inert filler, as well as indications of thermal stability

Test Procedure:
Set the inert (usually N2) and oxidative (O2) gas flow rates to provide the appropriate environments for the test. Place the test material in the specimen holder and raise the furnace. Set the initial weight reading to 100%, then initiate the heating program. The gas environment is preselected for either a thermal decomposition (inert - nitrogen gas), an oxidative decomposition (air or oxygen), or a thermal-oxidative combination.

Specimen size:
10 to 15 milligrams

Data:
A plot of percent weight loss versus temperature.

Equipment Used:
Perkin Elmer TGA7 Thermogravimetric Analyzer


Intertek PTL Overview
Testing
Quality
Contact Us
Services & Tech Info
Industry Segments
Industry Segments

Search PTLI.com

**Please note that this test description is intentionally generic in nature and aimed at providing a descriptive summary to enhance test understanding. For more information please contact a Intertek PTL Technical Representative at . Due to copyright restrictions, we are not able to provide copies of standards. Standards can be obtained from appropriate standards authorities.